Gas ballast, water vapour tolerance and why pumps drown

7 мин чтения

Bangkok air in the rainy season holds about 25 grams of water per cubic metre. A 300 m³/h vacuum pump evacuating a chamber of that air inhales 7.5 kg of water an hour before the process adds anything. If the pump is oil-sealed and the gas ballast is closed, a good share of that water ends up in the oil, and the pump is on its way to a rebuild. This note explains what is happening inside the pump, what the ballast does about it, and the figures and routines that keep an oil-sealed pump healthy in a humid plant.

1. Why vapour condenses inside a pump

A rotary vane pump draws a volume of gas in at the inlet pressure and compresses it, in the space between two vanes, until the exhaust valve opens at slightly above atmosphere. If the inlet is at 10 mbar, the compression ratio is about 100 : 1.

Water vapour has a saturation pressure that depends only on temperature: 23 mbar at 20 °C, 74 mbar at 40 °C, 312 mbar at 70 °C, 1,013 mbar at 100 °C. Compress a gas containing vapour and the vapour's partial pressure rises with the total pressure. The moment the partial pressure of water exceeds its saturation pressure at the pump's temperature, water condenses — in the compression chamber, into the oil.

Numbers: inlet at 10 mbar, gas is 50 % water vapour (a drying process), so the water partial pressure at the inlet is 5 mbar. Compressed 100 : 1 to open the exhaust, the water partial pressure would be 500 mbar. The pump runs at 70 °C, where saturation is 312 mbar. Condensation begins when the compression has reached 62 : 1 — a third of the way through the stroke the water starts to rain out, and every cycle leaves liquid water behind in the oil.

Why does the pump then lose vacuum rather than just get wet?

Because the oil is the seal, and water in the oil has a vapour pressure of its own. Oil contaminated with water forms an emulsion whose vapour pressure at pump temperature is tens of millibar — the ultimate pressure of the pump rises from 0.5 mbar to 5, 10, 20 mbar, because the pump is now also pumping the water out of its own oil. The vanes, running in emulsion rather than oil, lose lubrication and the pump gets louder and hotter. Corrosion starts on the rotor and stator. Within weeks the process "cannot reach vacuum", and the pump is blamed.

2. What the gas ballast does

The gas ballast is a small valve that admits atmospheric air into the compression chamber after the inlet has closed. That air adds to the total gas in the chamber, so the exhaust valve opens earlier in the stroke — at a lower compression ratio. The water vapour never reaches its saturation partial pressure, and leaves the pump as vapour in the exhaust.

With the ballast open:

  • The water partial pressure at exhaust is diluted; with a ballast flow of about 10 % of displacement, the pump tolerates an inlet water vapour partial pressure of roughly 30–40 mbar at 70 °C instead of about 3 mbar with the ballast closed.
  • The ultimate pressure rises slightly — a V140 goes from 0.5 mbar to about 1–2 mbar, a DV60 from 5 × 10⁻³ to 4 × 10⁻² mbar — because the ballast air is itself a gas load. For almost every industrial process this is irrelevant.
  • The pump runs a few degrees hotter, which helps.
  • Power rises slightly and the exhaust is louder.

The rule, then: ballast open whenever vapour is present, which in a Thai plant is always unless the process is bone-dry nitrogen. Close it only for the last decade of a high-vacuum pump-down on a dry, clean system, and open it again afterward.

3. Reading the figures

Datasheets give the water-vapour capability in two forms:

  • Water vapour tolerance — the maximum inlet partial pressure of water vapour (mbar) at which the pump will pump pure vapour without condensing, with ballast open. Typical: 30–40 mbar for single-stage vane pumps.
  • Water vapour capacity — the mass of water the pump can remove continuously, in kg/h, with ballast open. This is the practical number. Dr.Vakuum V Series figures at 50 Hz:
Pump Speed Water vapour capacity
V10 10 m³/h 0.5 kg/h
V63 63 m³/h 1 kg/h
V100 100 m³/h 1.5 kg/h
V140 132 m³/h 2 kg/h
V301 300 m³/h 4.5 kg/h
V630 640 m³/h 17 kg/h

Compare against the process. A chamber packaging machine on frozen product releases perhaps 0.2–0.5 kg/h — well inside the V301's figure. A tray dryer evaporating 20 kg/h is beyond any oil-sealed pump's capacity and needs a condenser upstream or a liquid ring pump; the ballast cannot rescue it. The capacity number is the one that tells you whether you are in ballast territory or condenser territory.

4. Routines that make the figure real

Three habits, all free:

  1. Warm up before vapour. The saturation pressure at 70 °C is fourteen times that at 20 °C. A pump started cold and fed vapour immediately condenses at a compression ratio it would sail through when hot. Run 15–20 minutes against a closed inlet (ballast open) before opening to a wet process. Where the process cannot wait, fit an oil heater or keep the pump idling.
  2. Ballast open, locked. On a packaging or drying pump, remove the temptation: wire the ballast valve open or specify a pump with a fixed ballast orifice.
  3. Purge at end of shift. Thirty minutes with the inlet closed and the ballast open boils absorbed water back out of the oil before it emulsifies. If the oil is already milky, change it — do not top up. An oil that stays milky after a purge has an emulsion that will not break; it is finished.

And one piece of hardware: an exhaust oil mist separator that drains back into the pump returns oil, but a separator on a wet duty also returns the condensed water. Fit a separator with a drain cock and empty it, or a coalescing element that discharges water separately.

5. When the honest answer is a different pump

  • Vapour load above the kg/h capacity → condenser ahead of the pump, sized to the real cooling water temperature (see Inlet filtration, condensers and knock-out pots), or a liquid ring pump (LM, LT) for which water is the working fluid.
  • Process below 1 mbar with vapour → the ballast's ultimate penalty starts to matter; a two-stage DV Series pump keeps a 4 × 10⁻² mbar ultimate with ballast and has a hotter second stage.
  • Solvent vapour → ballast works the same way, but the solvent dissolves in the oil and is not boiled off by a purge; a dry screw (DSI Series) with hot rotors passes the solvent and recovers it in a condenser.
  • Vapour plus oil-free requirement → dry claw (CL) passes condensate and drains it; dry vane (DRV) does not like continuous vapour and should be kept warm and dry.

6. Dry pumps drown too

The problem is not unique to oil. A dry vane pump condenses water in its chamber when cold and the carbon vanes swell; a dry screw pump's rotors run hot enough to pass vapour, but condensate in the exhaust line can run back into a stopped pump — the exhaust must fall away from the pump with a drain at the low point, and the pump should purge with nitrogen or run on dry air for a few minutes before stopping. A claw pump has the same exhaust-drain requirement. The physics of saturation pressure does not care what the pump is made of.

Recommended Dr.Vakuum machines for wet duties

Oil-sealed vane with gas ballast (V Series) — up to the kg/h figures in §3, ballast open, warm: V140, V301, V630.

Two-stage (DV Series) — wet processes below 1 mbar: DV90, DV175.

Liquid ring (LM / LT Series) — vapour loads beyond any ballast: LM200-065, LT230-160V.

Dry screw (DSI Series) — solvent vapour, recovered downstream: DSI300.

The failure described here is the one behind most warranty claims on packaging pumps; the companion note on vacuum packaging of frozen seafood and meat shows the routines applied to a real line.