Dental suction is the most common vacuum system in any Thai town — every surgery has one — and the most consistently undersized, because it is bought by the chair rather than by the practice. The high-volume evacuator at the chairside moves 250–300 litres of air per minute with a mix of water, saliva, blood and aerosol in it, at a modest vacuum of 150–200 mbar below atmosphere. Multiply by the number of chairs working at once and you have the plant. This note covers how to do that multiplication, what to do with the liquid, and why dry pumps have taken over.

1. What the system has to do

Two suction functions share one pipeline:

  • High-volume evacuation (HVE) — the large cannula the assistant holds. It removes the water spray from the handpiece and the aerosol from the mouth. It needs flow: about 250–300 L/min per cannula at the tip, at 150–200 mbar (15–20 kPa) of vacuum. More vacuum does not help; it pulls soft tissue into the tip.
  • Saliva ejector — the small low-volume tube. 30–60 L/min, same vacuum.

Design standards (ISO 10637 for dental suction equipment, and the general hospital standard ISO 7396-1 where the clinic is inside a hospital) call for the HVE flow to be available at the chair with the specified number of chairs in simultaneous use, after the pipeline losses, with the separator and filter in place.

2. Sizing by chairs and simultaneity

Not every chair is at the drilling stage at once. The accepted simultaneity factors:

Chairs Chairs in simultaneous HVE use Design flow at pump, free air
1 1 300 L/min (18 m³/h)
2–3 2 600 L/min (36 m³/h)
4–6 3 900 L/min (54 m³/h)
7–10 4–5 1,200–1,500 L/min (72–90 m³/h)
11–20 (polyclinic, faculty) 6–8 1,800–2,400 L/min (108–144 m³/h)
20+ (dental hospital) 30–40 % of chairs by calculation

The flow figures are free air at the pump inlet, and the pump must deliver them at the working vacuum, not at open inlet. A dry vane pump rated 48 m³/h at atmosphere delivers very nearly that at 200 mbar differential; a side channel blower rated 48 m³/h delivers perhaps half. Read the curve at 800 mbar absolute.

Pipeline losses are not trivial in a clinic with 40 m of 32 mm pipe: at 900 L/min the friction loss in that run is 30–50 mbar, so a plant set at 200 mbar gives 150–170 at the chair. Size pipework for velocity under 8 m/s — 32 mm to 3 chairs, 40 mm to 6, 50 mm beyond — and the plant set point at 220–250 mbar below atmosphere.

Why is more vacuum not the fix for weak suction?

Because the limiting element is the cannula tip and the hose, both of which are fixed small orifices. Flow through an orifice rises with the square root of the pressure difference and then chokes: beyond about 500 mbar differential the air speed at the tip reaches sonic and flow stops increasing at all. Doubling the vacuum from 200 to 400 mbar raises tip flow by perhaps 30 %, and causes tissue injury. Weak suction is almost always a flow shortfall from too many chairs on too small a pump, a partly blocked filter or separator, or a leaking pipeline — all of which are fixed by flow, not by depth.

3. Wet, dry and semi-dry systems

Wet systems draw air and liquid down the same pipe to a water-ring pump at the plant, which uses the liquid as its sealing water and discharges everything to drain. Simple, and standard for decades; but the pump consumes fresh water continuously (several litres a minute), the drain carries amalgam unless separated, and the pump is noisy and wet.

Dry systems separate the liquid from the air at each chair or at a central air-water separator ahead of the pump; the pump handles air only. The pump can then be a dry-running machine with no water consumption, lower noise and a sealed exhaust. Nearly all new installations are dry systems, and most wet-system replacements convert.

Semi-dry systems separate centrally with a single large separator and a dry pump; this is the usual layout for a clinic above three chairs.

Whichever system, the plant has four components in order: the amalgam separator (required in most jurisdictions, ISO 11143, ≥ 95 % retention), the air-water separator with a drain pump, a bacterial filter on the air side, and the vacuum pump with its exhaust piped outside. Add a non-return valve and a vacuum relief valve so a blocked line cannot stall the pump at full differential.

4. Choosing the dry pump

Three dry technologies serve the duty.

  • Dry rotary vane (DRV Series): oil-free carbon vanes, 90 mbar ultimate, quiet, compact. The right machine for 1–6 chairs. Vanes are a wear part — 5,000–10,000 hours, an afternoon's job to change.
  • Dry claw (CL Series): non-contact, no wear parts in the chamber, 150 mbar ultimate, higher flow. The right machine from 6 chairs up and for anything that must run all day — dental hospitals, faculty clinics.
  • Side channel blower (2SC Series): used by some package suppliers for very large plants; a two-stage blower holds 350–400 mbar, enough for a 220 mbar set point, and is dry and cheap. It is a less forgiving machine against a blocked filter — it overheats at closed inlet — and needs a relief valve.

The ultimate pressure of 90–150 mbar on the dry vane and claw pumps is a feature here: the set point is 220–250 mbar, so the pumps work in the flat part of their curve, and there is no way to over-vacuum a patient.

5. Worked plants

Three-chair private clinic, dry system. Design flow 600 L/min (36 m³/h) at 220 mbar. A DRV40 (48 m³/h, 2.2 kW) with a central air-water separator and amalgam unit, set point 230 mbar, 40 mm main. One pump; the clinic closes for pump maintenance twice a year.

Twelve-chair polyclinic. Design flow 1,800 L/min (108 m³/h) at 220 mbar. Two CL80 claw pumps (75 m³/h each) in duplex — either one carries 70 % of design, both together carry it with margin, and the second is also the standby that the clinic's accreditation audit asks about. Central separator sized for 150 L/min of liquid, bacterial filter duplexed, exhaust to roof.

Dental faculty, 60 chairs. Simultaneity 35 % → 21 chairs → 6,300 L/min (380 m³/h) at 220 mbar. Three CL150 (150 m³/h each) in triplex with lead/lag/standby rotation; or two CL251 duplex. The pipeline is a ring main in 80 mm with 50 mm drops to each clinic bay, and the plant sits in a dedicated room because 380 m³/h of exhaust is warm and audible.

6. Infection control and the exhaust

Dental suction air is aerosol-laden; since 2020 the handling of it has been scrutinised. The bacterial filter on the air side should be HEPA grade (≥ 99.97 % at 0.3 µm) and changed as clinical waste. The pump exhaust goes outside the building, away from windows and air intakes, with a drain at the low point for condensate. Dry pumps make this straightforward; wet systems discharge contaminated water to drain and aerosolise some of it at the pump.

Daily: run the line cleaner through every chair at close (the manufacturer's non-foaming disinfectant — foaming agents fill the separator). Weekly: check the separator drain pump and the filter differential. Annually: pump service and an HVE flow test at the furthest chair with the design number of chairs open.

Recommended Dr.Vakuum machines

Dry rotary vane (DRV Series) — single surgeries to six chairs, oil-free, quiet:

Model Speed, 50 Hz Motor Chairs (simultaneous)
DRV25 30 m³/h 1.5 kW 1–2 (1)
DRV40 48 m³/h 2.2 kW 2–3 (2)
DRV60 — — 4–6 (3)
DRV100 — — 7–10 (4–5)

Dry claw (CL Series) — polyclinics, faculties and hospitals, duplex or triplex: CL80 (75 m³/h), CL150 (150 m³/h), CL251 (250 m³/h).

For a clinic inside a hospital, the plant falls under the medical vacuum standard — see Hospital central medical vacuum to ISO 7396-1.