Laboratory vacuum: rotary evaporators, filtration and central lab vacuum lines
Every university chemistry department, every pharmaceutical QC laboratory, every food-testing and petrochemical analysis lab in Thailand runs on vacuum, mostly from small pumps under benches and in fume hoods, and increasingly from a central laboratory vacuum line piped to every bench like compressed air. The duties are small and the numbers are large: a 20-bench lab may have forty vacuum points. The choices — which pump, oil or not, bench or central — are made once per building and lived with for twenty years.
1. Three pressure levels
Laboratory vacuum divides cleanly by what the pressure is for:
| Level | Pressure | Duties | What matters |
|---|---|---|---|
| Rough | 100–300 mbar abs | Büchner and sintered-glass filtration, aspiration of liquids, vacuum desiccators, SPE manifolds, gel dryers | Flow (to pull air through a filter cake), tolerance of liquid and vapour, oil-free |
| Fine | 5–50 mbar abs | Rotary evaporators, vacuum ovens at 40–60 °C, concentrators, distillation under reduced pressure | Controllable pressure (to boil the solvent at the bath temperature), solvent vapour tolerance, cold trap |
| High | < 1 mbar (to 10⁻³) | Schlenk lines, freeze dryers, high-vacuum drying, molecular stills, GC/MS backing | Ultimate pressure, cleanliness (no back-streaming), two-stage |
The water aspirator (water-jet pump) served the first two levels in every lab for a century: 15–30 mbar, tolerant of anything, 10–20 litres of water a minute down the drain. Thai universities and laboratories are retiring them — a single aspirator running eight hours a day uses 5–10 m³ of water, and the solvents it pulls go to the drain with it. A diaphragm pump or a small dry vane pump replaces them.
Why does the rotary evaporator need a controlled pressure, not just a deep one?
Because the solvent boils at the pressure's boiling point, and the operator wants it to boil gently at the bath temperature, not flash and bump into the condenser. For ethanol at a 40 °C bath the right pressure is about 175 mbar; for water 72 mbar; for dichloromethane 700 mbar (barely below atmosphere); for DMF 10 mbar. A pump that pulls to 5 mbar regardless makes every solvent bump. Rotary evaporators therefore run with a vacuum controller — a solenoid valve or a variable-speed pump holding a set point — and the pump's job is to have enough speed to reach the set point quickly and enough ultimate to reach the lowest set point the lab uses (DMF, DMSO at 5–10 mbar). A 90 mbar dry vane pump is wrong for a rotary evaporator; a 0.5 mbar oil-sealed or a diaphragm pump reaching 2–7 mbar is right.
2. Bench pumps
For the rough level: oil-free dry vane pumps. A DRV10 (12 m³/h, 90 mbar ultimate) serves a filtration station or a 4-port SPE manifold; the gas carries water and solvent droplets, so a trap flask with a float valve between glassware and pump, and the pump's exhaust to the fume hood duct. Diaphragm pumps (not in our range) cover the same duty at lower flows.
For the fine level: either a chemistry diaphragm pump or an oil-sealed rotary vane pump with a cold trap. The oil-sealed pump — a V10 (10 m³/h, 2 mbar) or V21 (20 m³/h, 1 mbar) — is cheaper, reaches lower pressure and has more speed, which is why rotary evaporator benches in Thai universities still mostly run on them. The price is the oil: solvent vapour dissolves in it, the ultimate rises, and the oil must be changed monthly on a busy evaporator. A cold trap (dry ice/acetone or a −80 °C refrigerated trap) between evaporator and pump catches most of the solvent, and gas ballast open handles the rest — see Gas ballast, water vapour tolerance and why pumps drown. A pump dedicated to one evaporator with a controller is the standard good practice.
For the high level: two-stage oil-sealed — DV5 (5.4 m³/h, 5 × 10⁻⁴ mbar) for a Schlenk line or a small freeze dryer, DV16 (14 m³/h, 5 × 10⁻³) for a vacuum oven or a GC/MS backing duty — always with a cold trap, and with a foreline trap where the instrument is hydrocarbon-sensitive.
3. A central laboratory vacuum line
Above about ten vacuum points, a central system — a duplex pump set in a plant room, a receiver, and a piped network with a valve at each bench — is cheaper to install than bench pumps, far quieter, and removes oil and exhaust from the lab. It serves the rough level only: the line is held at 100–200 mbar absolute (−800 to −900 mbar gauge), which suits filtration, aspiration, desiccators and SPE. Rotary evaporators and high-vacuum work keep their own pumps, because they need controlled or deeper pressure.
Sizing follows the medical-vacuum method — points, flow per point, diversity:
- Each bench outlet is rated at 30–60 L/min of free air (a Büchner funnel pulls air through the cake continuously).
- Diversity: 20–30 % of points in use at once in a teaching lab, 40–50 % in a busy QC lab.
- A 20-bench QC lab with 40 points at 40 L/min and 50 % diversity: 800 L/min = 48 m³/h; the pumps must deliver this at 150 mbar, with the standard requirement of duplex capacity (either pump carries the load).
Two DRV40 (48 m³/h each, 90 mbar ultimate) in duplex with a 300 L receiver and a pressure switch — oil-free, so the exhaust from the plant room needs only a particle filter. For a research building with 150 points, two or three CL80 claw pumps.
The network:
- Pipe in copper or stainless (not PVC — solvent vapour), sloping back to the plant room, 22 mm to a bench group, 42–54 mm risers.
- A trap flask or in-line separator at every outlet — the central line must never receive liquid; a flooded riser is a plant-room problem.
- A central knock-out pot with a drain and a bacterial / solvent filter before the pumps.
- Non-return valve per floor, so one lab's flood does not pull on another.
- Pressure gauge and alarm on the receiver; alarm at −600 mbar.
What the central line must not carry: concentrated solvent vapour (rotary evaporators stay off it), corrosive gases (HCl, Cl₂ digestions — local scrubbed pumps), anything from a biosafety cabinet (dedicated HEPA-filtered pump), and radioactive work.
4. Oil-free or oil-sealed, for a laboratory
| Oil-free dry vane / diaphragm | Oil-sealed rotary vane | |
|---|---|---|
| Ultimate | 90 mbar (dry vane); 2–7 mbar (chemistry diaphragm) | 0.5–2 mbar single-stage; 5 × 10⁻³ two-stage |
| Solvent vapour | Passes through; corrodes diaphragm pumps slowly | Dissolves in oil; cold trap and ballast essential |
| Maintenance | Vanes 5–10 kh; diaphragms 10–15 kh | Oil change 1–3 months on solvent duty |
| Noise | Moderate | Low |
| Exhaust | Dry; duct to hood | Oil mist; separator and duct |
| Cost per m³/h | High (diaphragm), low (dry vane) | Low |
| Right for | Filtration, central lines, teaching labs | Rotary evaporators, ovens, high vacuum |
A sensible laboratory therefore runs dry on the central line and the benches, oil-sealed with traps on the evaporators and ovens, and two-stage oil-sealed with traps on the high-vacuum work.
5. Worked laboratories
| Laboratory | Points | Dr.Vakuum answer |
|---|---|---|
| Teaching lab, 24 students, filtration only | 24 at 30 L/min, 30 % | Central: 2 × DRV25 duplex, 200 L receiver |
| Pharma QC lab, 20 benches | 40 at 40 L/min, 50 %; 6 rotary evaporators; 2 vacuum ovens | Central: 2 × DRV40; evaporators: 6 × V10 with controllers and cold traps; ovens: 2 × DV16 |
| Natural-products research, 8 hoods | Per hood | 8 × V21 with cold traps; 2 × DV5 Schlenk lines |
| Research building, 150 points, 5 floors | 150 at 40 L/min, 35 % → 126 m³/h at 150 mbar | Central: 3 × CL80 triplex, 1,000 L receiver, floor NRVs |
| Food testing lab, Kjeldahl and digestion | Corrosive | Local PTFE diaphragm pumps with scrubbers (not in our range); general filtration from a DRV10 |
Recommended Dr.Vakuum machines
Dry rotary vane (DRV Series) — filtration, aspiration, central lab lines, oil-free: DRV10 (12 m³/h), DRV25 (30 m³/h), DRV40 (48 m³/h).
Dry claw (CL Series) — large central systems: CL80.
Single-stage oil-sealed vane (V Series) — rotary evaporators and ovens with cold trap and controller: V10 (10 m³/h, 2 mbar), V21 (20 m³/h, 1 mbar).
Two-stage oil-sealed vane (DV Series) — Schlenk lines, freeze dryers, instrument backing: DV5, DV16.
The hospital version of the central line, with its stricter simultaneity and standards, is in Hospital central medical vacuum to ISO 7396-1.
Lab to schedule. Send us the floor plan with vacuum points, the number of evaporators and ovens, and the solvents in use, and we will return a central-line size and a bench-pump list. Talk to an engineer.